Shopping Cart
Shop Now
Login
Search for products on our site
See more
BLT M9000Vet Biolight with CO2 Veterinary Monitor
BLT M9000-Biolight with CO2 Veterinary Monitor
BLT M9000Vet Biolight with CO2 Veterinary Monitor
BLT M9000Vet Biolight with CO2 Veterinary Monitor
On Sale

BLT M9000Vet Biolight with CO2 Veterinary Monitor

No reviews
Regular price $3,399.00

Biolight BLT M9000 Capnograph Patient Monitor with CO2 side stream.  Suitable for adult, pediatric and neonateIt.  It includes 7 parameters (ECG, RESP, NIBP, SPO2, TEMP, IBP, EtCO2).  Has large 12.4" screen LCD.  Features of the monitor are listed below.

  • Safety: IEC60601 (GB9706)
  • Dimension: 318mm (W) x 264mm (H) x 152mm(D)
  • Power Supply : 100~240V; AC: 50~ 60 Hz
  • Alarm Indication: Visual and audible alarm
  • 1000 groups of alarm events review
  • Application: Adult, Pediatric and neonate

 ECG 

  • Leads: I, II, III, aVR, aVL, aVF, V
  • Gains: 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, auto
  • Input impedance: = 5MO (d.c.~60Hz)
  • Common Mode Rejection Ratio: = 85dB (diagnostic mode) =105dB (monitoring & surgical mode)
  • Bandwith: 0.5-40Hz (monitoring mode) 0.05-120Hz (diagnostic mode) 1-25Hz (surgical mode)
  • Protection: Segregation breakdown voltage 4000VAC, 50/60Hz Resistant to high-frequency electrotome, myoelectricity, AC, drifting interference
  • Indication of Detachment of electrodes: Available
  • Playback of Cardiogram: Available
  • Scanning Speed: 12.5mm/s, 25mm/s, 50mm/s
  • Heart Rate Range: 10-300BPM
  • Accuracy: ±1BPM
  • ECG waveform playback: 10 minutes

NIBP 

  • Method: Automatic oscillometric
  • Operation Modes: Singles (manual), Cycle (automatic), Rapid (continuous)
  • Measurement : Systolic pressure/diastolic pressure/mean pressure
  • Range: Adult: 10-270mmHg
  • Pediatric: 10-235mmHg
  • Neonate: 10-135mmHg
  • Unit: mmHg / Kpa selectable
  • Auto Measurement: 1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90 minutes, 2, 4, 8, 12 hours
  • 750 groups of NIBP measuring data review

RESP 

  • Method: Impedance Measurement
  • Measuring Range: 0-120bpm
  • Alarm Limit: 0-120bpm, upper and lower limit continuously adjustable
  • Apnea alarm: Available
  • BLT Digital SPO2 (Pulse Oximetry)
  • Range: 0-100%
  • Accuracy: ±2% (70%~100%); ±3%(35%~ 69%) Unspecified (0~34%)
  • Range & Accuracy of Pulse Rate : 20~250BPM ± 1BPM

TEMP 

  • Channel : 2 channels
  • Range : 0~50°
  • Unit : Celsius, Fahrenheit
  • Resolution : 0.1
  • Accuracy : ±0.1° (25°~50°)
  • Alarm Range : 0~50°

 IBP 

  • Channel : 2 channels
  • Range : -50~300 mmHg
  • Resolution : 1 mmHg
  • Accuracy : ±2 mmHg (±2%) without probe ±4 mmHg (±2%) with probe
  • Unit : mmHg , Kpa
  • Converter sensitivity: 5 uV/V/mmHg
  • Names of pressures: ART/PA/CVP/LAP/RAP/ICP

EtCO2 (side stream) 

  • Method: Infrared spectrum
  • Range: 0.0-13.1%
  • Resolution : 1 mmHg (0.1%)
  • Accuracy : 5% EtCO2