    Biolight BLT M9000 Capnograph Patient Monitor with CO2 side stream.  Suitable for adult, pediatric and neonateIt.  It includes 7 parameters (ECG, RESP, NIBP, SPO2, TEMP, IBP, EtCO2).  Has large 12.4" screen LCD.  Features of the monitor are listed below.

    • Safety: IEC60601 (GB9706)
    • Dimension: 318mm (W) x 264mm (H) x 152mm(D)
    • Power Supply : 100~240V; AC: 50~ 60 Hz
    • Alarm Indication: Visual and audible alarm
    • 1000 groups of alarm events review
    • Application: Adult, Pediatric and neonate

     ECG 

    • Leads: I, II, III, aVR, aVL, aVF, V
    • Gains: 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, auto
    • Input impedance: = 5MO (d.c.~60Hz)
    • Common Mode Rejection Ratio: = 85dB (diagnostic mode) =105dB (monitoring & surgical mode)
    • Bandwith: 0.5-40Hz (monitoring mode) 0.05-120Hz (diagnostic mode) 1-25Hz (surgical mode)
    • Protection: Segregation breakdown voltage 4000VAC, 50/60Hz Resistant to high-frequency electrotome, myoelectricity, AC, drifting interference
    • Indication of Detachment of electrodes: Available
    • Playback of Cardiogram: Available
    • Scanning Speed: 12.5mm/s, 25mm/s, 50mm/s
    • Heart Rate Range: 10-300BPM
    • Accuracy: ±1BPM
    • ECG waveform playback: 10 minutes

    NIBP 

    • Method: Automatic oscillometric
    • Operation Modes: Singles (manual), Cycle (automatic), Rapid (continuous)
    • Measurement : Systolic pressure/diastolic pressure/mean pressure
    • Range: Adult: 10-270mmHg
    • Pediatric: 10-235mmHg
    • Neonate: 10-135mmHg
    • Unit: mmHg / Kpa selectable
    • Auto Measurement: 1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60, 90 minutes, 2, 4, 8, 12 hours
    • 750 groups of NIBP measuring data review

    RESP 

    • Method: Impedance Measurement
    • Measuring Range: 0-120bpm
    • Alarm Limit: 0-120bpm, upper and lower limit continuously adjustable
    • Apnea alarm: Available
    • BLT Digital SPO2 (Pulse Oximetry)
    • Range: 0-100%
    • Accuracy: ±2% (70%~100%); ±3%(35%~ 69%) Unspecified (0~34%)
    • Range & Accuracy of Pulse Rate : 20~250BPM ± 1BPM

    TEMP 

    • Channel : 2 channels
    • Range : 0~50°
    • Unit : Celsius, Fahrenheit
    • Resolution : 0.1
    • Accuracy : ±0.1° (25°~50°)
    • Alarm Range : 0~50°

     IBP 

    • Channel : 2 channels
    • Range : -50~300 mmHg
    • Resolution : 1 mmHg
    • Accuracy : ±2 mmHg (±2%) without probe ±4 mmHg (±2%) with probe
    • Unit : mmHg , Kpa
    • Converter sensitivity: 5 uV/V/mmHg
    • Names of pressures: ART/PA/CVP/LAP/RAP/ICP

    EtCO2 (side stream) 

    • Method: Infrared spectrum
    • Range: 0.0-13.1%
    • Resolution : 1 mmHg (0.1%)
    • Accuracy : 5% EtCO2